Optics Characterization Tests
UUID: F66D66DC-23E9-42EC-A635-1D8BE9832C16
Optics characterization should produce test artifacts for:
Wavelength accuracy after pixel-to-wavelength calibration.
FWHM of narrow emission lines.
SNR at representative visible and near-infrared wavelengths.
Linearity versus input intensity and integration time.
Repeatability after repeated captures and after minor handling.
Stray light and second-order diffraction behavior.
Comparison with a reference source or commercial spectrometer.
The test record should include optical configuration, slit width, grating setting, filter state, integration time, calibration version, and raw exported data path.