ADC And DMA Capture
UUID: 8F4E69AC-05DC-4E3C-8CF5-4E84DA7BBE2A
The acquisition path shall sample the CCD analog output from timer-triggered ADC conversions and use DMA to move samples into frame storage with bounded CPU intervention.
Effective detector content: Each accepted capture shall preserve all 3648
effective detector pixels in their acquisition order. The requirement does
not currently define the number or placement of non-effective samples, so the
total transported sample count, dummy regions, sample width, byte order, ADC
trigger frequency, sample phase, input range, clipping behavior, and buffer
ownership are Not In Docs open requirements.
Frame completion: A frame is eligible for transfer only after the required effective samples have been acquired and the controller has recorded whether the capture completed normally. A partial or invalid capture shall be identified as such rather than silently padded or presented as a complete measurement.
Requirement status: documented intended behavior; ADC/DMA implementation
alignment is Not Started.